Oxygen Ratio Controller (ORC): In the ORC a slave valve pneumatically linked to two pressure balancing diaphragms of Oxygen & guarantees a minimum of 25% Oxygen Concentration which can be raised anywhere in the range of 25% to 100% (±5%.)
Oxygen Failure Protection Device (OFPD): MEDISYS OFPD works as a true Fail-safe system allowing flow of only in presence of Oxygen at Rotameter level since it locks the flow of in the absence of flow of Oxygen through its Rotameter.
Basal Flow: When MEDISYS OFPD is in operation as indicated by Audio-Visual Alarm, a separate Pneumatic Circuit is activated to deliver a minimum Oxygen Flow of 200 - 300 ml/min through its Rotameter.
Alarm: Audio-Visual Alarm is activated if flow of Oxygen falls below LPM.
Pneumatic Circuitry: Completely encased, detachable Pneumatic Circuitry for easy & cprompt service check up.
Circle Absorber System: Twin Canister Circle Absorber System is provided with swivelling facility. Canisters and Inspiratory Valve have quick -fit Taper Fitting with Rubber '0' Rings for Leak proof seal and easy maneuverability. Bag-Mount & APL Valve Assembly have swivel arms for easy accessibility.
Emergency Oxygen: Emergency Oxygen Flush Button is provided at Table Top level on the front. It is recessed and guarded to avoid accidental pressing. Flow Range: 35 LPM Calibrated at 60 PSI (±5%).
Vaporizers: The Vaporizers of Ether and Halothane are supplied as standard accessories. Provision for incorporating Vaporizer of user's choice in the Back Bar, Viz. Floutec, Penlon etc. (Optional).
Non-Return cum Pressure Relief Valve: Blows off excess pressure in the Circuit. Release Pressure: 100 CM H20 (±10%.).